Please use this identifier to cite or link to this item: https://hdl.handle.net/10316.2/44455
DC FieldValueLanguage
dc.contributor.authorBravim, A.
dc.contributor.authorSakuraba, R. K.
dc.contributor.authorCampos, L. L.
dc.date.accessioned2018-11-05T16:52:08Z
dc.date.accessioned2020-09-05T07:51:53Z-
dc.date.available2018-11-05T16:52:08Z
dc.date.available2020-09-05T07:51:53Z-
dc.date.issued2018-
dc.identifier.isbn978-989-26-1602-5 (PDF)
dc.identifier.isbn978-989-26-1601-8
dc.identifier.urihttps://hdl.handle.net/10316.2/44455-
dc.description.abstractThe objective of this study was to evaluate the applicability of OSL dosimeters and microdosimeters of LiF: Mg, Ti (TLD-100) in determining isodose curves in radiotherapy planning using the VMAT technique. A vertebra treatment plan with spinal cord protection was simulated; the mean absorbed dose obtained through the OSL dosimeters presented a maximum variation of 3.85% in relation to the values calculated by the Eclipse 11.0 planning system. The variation between the dose range obtained using the OSL dosimeters and provided by Eclipse 11.0 agrees with acceptable limits for the radiotherapy area. It was observed that because of their smaller size, LiF: Mg, Ti microdosimeters are less likely to be positioned in more than one isodose curve at the same time in the planning. Therefore, they presented the best results.eng
dc.description.abstractO objetivo deste trabalho foi avaliar a aplicabilidade de dosímetros e microdosimetros OSL de LiF:Mg,Ti (TLD-100) na determinação das curvas de isodose em planejamentos utilizando a técnica de VMAT. Foi simulado um planejamento de tratamento de vértebra com proteção da medula espinhal em que a dose média absorvida obtida através dos dosímetros OSL apresentou uma variação máxima de 3,85% em relação aos valores calculados pelo sistema de planejamento Eclipse 11.0. A variação entre o intervalo de dose obtido com o emprego dos dosímetros OSL e fornecido pelo Eclipse 11.0 está de acordo com os limites aceitáveis para a área da radioterapia. Observou-se que, por serem de menor tamanho, os microdosímetros de LiF:Mg,Ti apresentam uma menor probabilidade de estarem posicionados em mais de uma curva de isodose ao mesmo tempo no planejamento. Sendo assim, apresentaram os melhores resultados.por
dc.language.isopor-
dc.publisherImprensa da Universidade de Coimbrapor
dc.relation.ispartofhttp://hdl.handle.net/10316.2/44443por
dc.rightsopen access-
dc.subjectVMATeng
dc.subjectisodose curveseng
dc.subjectoptically stimulated luminescenceeng
dc.subjectlithium fluorideeng
dc.subjectdosimetryeng
dc.subjectVMATpor
dc.subjectcurvas de isodosepor
dc.subjectluminescência opticamente estimulada,por
dc.subjectfluoreto de lítiopor
dc.subjectdosimetriapor
dc.titleAvaliação de dosímetros OSL na determinação de curvas de isodose de uma simulação de planejamento de tratamento de vértebra com proteção da medula espinhal utilizando VMATpor
dc.title.alternativeEvaluation of OSL dosimeters in determining isodose curves of a vertebra treatment planning simulation with spinal cord protection using VMATeng
dc.typebookPartpor
uc.publication.firstPage111-
uc.publication.lastPage120-
uc.publication.locationCoimbrapor
dc.identifier.doi10.14195/978-989-26-1602-5_12-
uc.publication.sectionCapítulo 2 - Dosimetria e instrumentaçãopor
uc.publication.digCollectionPBpor
uc.publication.orderno12-
uc.publication.areaCiências Naturaispor
uc.publication.bookTitleProteção contra radiações na comunidade dos países de língua portuguesa-
uc.publication.manifesthttps://dl.uc.pt/json/iiif/10316.2/44455/201062/manifest?manifest=/json/iiif/10316.2/44455/201062/manifest-
uc.publication.thumbnailhttps://dl.uc.pt/retrieve/11017274-
uc.publication.parentItemId55071-
uc.itemId68215-
item.fulltextWith Fulltext-
item.grantfulltextopen-
Appears in Collections:Proteção contra radiações na comunidade dos países de língua portuguesa
Files in This Item:
File Description SizeFormat 
avalicao_de_dosimetros_osl.pdf474.32 kBAdobe PDFThumbnail
  
See online
Show simple item record

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.